For ILR checks we use the abbreviated PBOP framework: Presenting rhythm, Battery status, Observations, Programming. The ILR is a monitor, not a therapeutic device — there are no leads, no pacing, and no therapy to test, so the L/S/T letters of the device-side PBL-STOP framework do not apply. Sensing checks (R-wave amplitude, artifact screen) live at the top of O because they’re part of what we’re observing, not a separate calibration step.
Device overview
An implantable loop recorder is a small subcutaneous monitor placed in the left parasternal area. It records a single-channel ECG continuously and stores episodes triggered automatically by algorithm or by the patient with a handheld activator or smartphone app. Most ILR follow-up is remote, with in-office visits reserved for specific issues — sensing problems, episode review for borderline findings, or end-of-indication decisions. Indications are narrow but important: unexplained syncope, palpitations without documented arrhythmia, cryptogenic stroke (AF surveillance), and post-ablation AF monitoring.
PBOP walkthrough
P — Presenting Rhythm
- Live strip at the start of the session — sinus vs AF vs other
- Snapshot heart rate and any obvious ectopy
- Symptoms in the room — palpitations, lightheadedness, recent syncope
- The clinical question driving the visit: are we still hunting for the same arrhythmia, or has the picture changed?
A 30-second chart pass before opening the episode list keeps the review focused.
B — Battery Status
- Voltage and estimated remaining longevity
- Expected total service typically 3–4.5 years depending on platform and episode volume
- Plan ahead:
- >12 months remaining: continue monitoring per indication
- 6–12 months: confirm whether the clinical indication still exists
- <6 months: plan explant vs replacement based on whether the diagnostic question is unresolved
Battery alone rarely drives a replace-vs-explant decision; the clinical question does.
O — Observations
The main event of an ILR check. Everything else exists to support this section.
Signal-quality snapshot first — before opening the episode list, confirm the device is recording cleanly. A 30-second baseline strip is enough.
- R-wave amplitude: should be >0.3 mV; trend it across visits — declining sensing is the leading cause of new false positives
- Sustained R-wave <0.2 mV degrades every downstream algorithm — AF, pause, brady, tachy
- Far-field and T-wave oversensing — double-counting that fakes tachycardia
- Myopotentials during arm/torso movement — common source of noise-driven false positives
- Body-position effects (supine vs upright) on platforms that expose them
- If sensing has dropped meaningfully, consider device migration, scar maturation, body habitus change, or a sensing-gain reprogramming
Triage order
- Patient-activated symptomatic episodes first — tied to symptoms, highest diagnostic value
- Then auto-detected episodes by clinical severity:
- Asystole / long pause
- AF
- Sustained tachycardia
- Bradycardia
- Non-sustained tachycardia
- HR-trigger episodes (rate-based, not rhythm-based)
EGM review for every flagged episode
- AF claims: irregular RR with no organised P-waves vs noise vs SVT vs frequent PACs
- Pause/asystole claims: true asystole vs R-wave undersensing vs lead artifact vs T-wave dropout
- Tachycardia claims: sinus vs SVT vs VT — width matters, onset matters
- Bradycardia claims: true sinus brady vs sensed pause from undersensing
- False-positive sources to know on sight: noise, myopotentials, PAC runs that mimic AF, sensing dropout that mimics pauses, T-wave oversensing that mimics tachycardia
Symptom correlation
- Cross-reference each device episode with the patient diary
- Symptomatic episode + documented arrhythmia = diagnostic
- Symptomatic episode + sinus rhythm = also diagnostic (rules out arrhythmic cause)
- Asymptomatic AF in a cryptogenic stroke patient = anticoagulation discussion
Memory management
- Clear reviewed episodes to free storage
- High event volume between transmissions can overflow memory and lose data
- If overflow is happening, tighten triggers or shorten the transmission interval
P — Programming
- Detection trigger adjustments: tighten if false positives are overwhelming the workflow; loosen if the clinical question demands maximum sensitivity (active cryptogenic stroke workup)
- Save EGM strips of any clinically relevant episode to the EMR
- Clear reviewed episodes per institutional policy
- Patient education refresh: activator use, what to do if symptomatic, when to call
- Save and print the summary report for the chart
- Set next remote transmission and next in-office visit
- Escalate to the EP attending for:
- Asystole >6 seconds, regardless of symptoms
- AF with stroke risk and no anticoagulation
- Any wide-complex tachycardia
- Syncope with a corresponding device-recorded arrhythmia
- Worsening symptoms with no device correlation — may need a different monitoring strategy
- End-of-indication explant decisions
A note on volume
The biggest workflow challenge with ILRs is volume. A single patient with a borderline AF algorithm can generate dozens of false positives per week. Triage policies — who reviews what, what gets escalated, and how often we revisit detection triggers — keep the device clinic functional. A trained ILR reader who knows the difference between noise, pseudo-AF from PACs, and true AF saves the practice an enormous number of hours.
Reference
Framework reference: “Keeping the Pace using PBL-STOP” — Chart Healthcare Academy